EmissionSEM_Picture1

Installed in late 2014 with an Oxford X-Max 50 Silicon 50mm2 drift detector (SDD) and a Gatan MonoCL4 imaging. The instrument is specifically configured for large-area imaging with digiscan software which allows for automated large area stitching of SE, BSE, EDS, and CL images. In addition the system is fitted with a “InLens” detector “for 3-4 nm image resolution under optimal conditions and two independent Cl modes (monochromatic, panchromatic) for imaging.

Detectors

EmissionSEM_Picture2

In AsB, grayscale is related to the atomic number (the lighter the higher the Z). Image of barite-filled vein in sandstone.

AsB (Angle-sensitive Backscatter) detector

This detector is mounted in the pole piece. The number of backscattered electrons produced in an elastic collision between electrons and particles (atoms) depends on the atomic number(Z). BSE provides Z-contrast, channeling contrast (crystallographic and strain information), and qualitative compositional analysis.

EmissionSEM_Picture3

Split-screen image showing the SE1 signal (right) and SE2 signal (left) of the same area.

Secondary electron (SE)

InLens detector (SE1) Secondary electrons are emitted from the top 10nm of the sample surface by initial interaction with the sample. Therefore, this detector provides surface details at low energy. This detector is designed to work at a range of beam energies, between 0.02-20 kV.

Everhart-Thornley detector (SE2) Secondary electrons are emitted after interacting with the topmost layer of the sample. This detector is designed to work at working distances between 5-12 mm and a range of beam energies, between 0.02-20 kV.

ImageGroup1_Picture1

X-ray energy-dispersive spectroscopy (EDS)

The EDS detector is used to separate the characteristic X-rays emitted from the excited sample into an energy spectrum. From this spectrum, the chemical composition of materials can be determined as well as the abundance of specific elements. Image: EDS map showing different elements in different colors.

ImageGroup1_Picture2

Variable-pressure secondary electron detector (VPSE)

This detector collects photons that are produced when a gas molecule is ionized, therefore it is designed specifically to work in a gaseous environment. VPSE imaging shows how different areas of a sample dissipate charge. It detects defects in a similar way to CL, but the signal is generated from a smaller interaction volume (higher resolution). Image: Whole-rock image of sandstone cataclasite outcrop plug.

ImageGroup1_Picture3

Monochromatic CL spectra

Our MonoCL4 detector can be set to spectrum-imaging mode. Each material emits a characteristic spectrum that can be used to distinguish between materials. The intensity of the photons emitted at each wavelength is recorded. Image: comparison between CL emission spectra within a deformation band and adjacent undeformed host rock.

Cathodoluminescence (CL) MonoCL4 detector

CL signal is produced when atoms bombarded by electrons return to the ground state and emit photons. CL signal depends on composition, lattice structure, and lattice damage. CL is useful for discerning cements and microfractures. Image: panchromatic CL image of a zoned zircon.

ImageGroup1_Picture4
ImageGroup1_Picture5

Color CL

Color CL images can be constructed by combining three images of the same area collected using a blue, green, and red filter.

ImageGroup1_Picture6
ImageGroup1_Picture7
ImageGroup1_Picture8
ImageGroup1_Picture9

Large-area mapping (SE1, SE2, AsB, VPSE, CL, and EDS)

Large-area mapping enables the collection of high resolution images automatically. Aztec software (Oxford Instruments) is used for large-area EDS mapping, whereas Digiscan (Gatan) is used for CL images collected with the MonoCL4 detector. Either one of the two softwares can be used for large-area mapping of the Zeiss detectors (SE1, SE2, AsB, VPSE). Simultaneous maps can be collected in a single run.

ImageGroup2_Picture1

Panchromatic CL image of a fractured tight-gas-sandstone composed of 1840 images. Stitching is automatic using Digiscan software. Inset of imaged area is composed of 12 tiles. We can routinely acquire images with a resolution of 10 nm per pixel.

ImageGroup2_Picture2

Color-CL mosaic composed of 49 images of a deformation band in sandstone. Stitching and image superposition is performed in Digiscan.

ImageGroup2_Picture3

Seamless EDS mosaic of the same area displayed in the image to the left. EDS mosaic composed of 20 images. Automated image acquisition and seamless stitching performed by Aztec.